注射器微量泵(USA)-精确可编程序-特价5折
- 来源:生物谷
- 时间: 2009-1-30
- 浏览人数: 1845
- Model : NE-1000 NE-1600 NE-1800
- Syringe sizes: 1 ~ 60 cc 1 ~ 60 cc 1 ~ 10 cc
- Number of syringes: 1~ 6 8
- Electrical
- Motor type: Step motor
- Motor steps per revolution: 200
- Motor to drive screw ratio: 1/4
- Drive screw pitch: 24 revolutions
- Advance per motor step: 1.323 um
- DC connector: 2.5 mm, center positive
- Voltage at DC connector: 12V DC at full load
- Amperage: 1000 mA at full load
- Power supply output rating: 12V DC @ 1000 mA
- Dimensions:
- NE-1000: 22.86 cm x 14.605 cm x 11.43 cm
- NE-1600,1800: 26.35 cm x 38.42 cm x 13.34 cm
- Weight: NE-1000 (1.63 kg) NE-1600,1800 (4.6 kg)
- Operational
- Maximum speed: 3.2226 cm/min
- Minimum speed: 0.0262 cm/hr
- pumping rate:
- Maximum force: 160 lbs at minimum speed
- Number of Program Phases: 41
- RS-232 pump network: 100 pumps maximum
- Syringe inside diameter range: 0.100 ~ 50.00 mm
- Infuses and withdraws
- Selectable pump rate units: µl/hr, µl/min, ml/hr, ml/min.
- Change pumping rate and direction while pumping.
- Dispense a specified volume.
- Syringe purge mode will infuse or refill syringe at top speed to purge air from the line or to fill the syringe.
- Audible buzzer can be programmed to alert when an alarm condition occurs or the pumping program completes.
- Power failure mode restarts a pumping program interrupted by a power failure.
- Stall detection stops pump when the motor's operation is impeded (not on NE-500 model)
- Non-volatile memory restores all setup parameters and the pumping program, on power up.
- Elastomer raised keypad with optional beeping whenever the keypad is pressed.
- CE certified
- Pre-program a set of production dispensing protocols. The operator then selects the required dispense protocol
- Keypad Lockout: Prevent unauthorized or accidental changes to the settings by requiring a "Lockout Disable" key to be inserted to change settings.
- 41 programmable phases
- Enter programs from the keypad or from a computer.
- Download or upload pumping programs from or to a computer. Store dispensing programs in a file on your computer, then download them to one or more pumps. Specify in the file to send a different pumping program to each pump in the pump network.
- Pre-program dispense volumes. Each volume can be different.
- Automatically change pumping rates or pumping direction.
- Ramp up or down the pumping rate.
- Set timed delays between dispenses.
- Automatically pause the program and wait for the user to continue the dispense.
- Synchronize dispenses with other equipment or pumps to change pumping rates in reaction to a sensor or signals from other pumps.
- Send logic signals to other equipment. Change pumping rates in reaction to a sensor.
- Program the audible alarm to beep at any time to alert the operator.
- Network up to 100 pumps from a single communications port on a computer. Baud rates up to 19200.
- Free terminal emulator program allows communications with syringe pump.
- Directly control one or more pumps from a computer, including setting the pumping rate, direction and volume. Set the operational and setup configurations. Also the TTL interface can be queried and the programmable output set.
- Using a computer, each pump in the pump network can be individually controlled and monitored. Or, the computer can download a new pumping program to each pump which would automatically reconfigure the dispensing system for a new job.
- Downloading control programs to the pump would unload computer overhead to the pump. Computer resources could then be used elsewhere while the pumps operate independently.
- 2 modes of communications: Basic and Safe
- Basic: Basic master-slave interface. Easy to implement. Can communicate from any terminal emulator.
- Safe: A reliable interface for a real-world environment.
- - High quality interface which detects communication errors via a 16 bit CCITT CRC.
- - The pump will automatically transmit an alarm packet when a pump alarm occurs. For instance, if the pump stalls or resets, a message is immediately sent to the computer.
- - A configurable communications time-out will stop the pump and alarm if the communications link is broken.
- - Packet inter-character time-out to recover packet synchronization after communication glitches.
- Software Developer friendly communications protocol.
- Consistent and well defined communications packets.
- Constant start and end characters.
- Packets structured to be easily parsed
- Control the pump from a foot switch or timer.
- Output signals can be used to switch a valve from an output to an input reservoir.
- Keypad Lockout: Prevent unauthorized or accidental changes to the settings by requiring a "Lockout Disable" key to be inserted to change settings.
- Pumping direction control
- Programmable output signal
- Programmable responses to input signals
- Configurable start and stop trigger:
- Start or stop the pump on each contact closure to ground. For example, a foot switch.
- Start the pump on closure to ground and stop the pump on opening the connection. For example a timer relay that starts the pump whenever closed and stops the pump whenever open.
- Only start the pump on each contact closure to ground. For example, a laboratory animal trained to press a lever can receive a drug dosage from the pump and the pump will not stop or deliver another dosage until it is permitted by the pumping program entered by the user
- Configurable trigger input:
- Edge trigger: Starts or stops the pump, such as from a foot switch
- Level trigger: Start or stops the pump, such as from a timer or relay control
- Start only trigger: Edge trigger will only start the pump, such as from a lab animal trained to press a lever to receive a dosage.
- Heating device for syringes that require temperature-controlled dispensing
- Heats syringe to over 100 degrees Celsius
- Digitally set the heater setpoint
- Temperature sensor in heating pad
- Setpoint retained in memory
- Controller will heat the syringe and hold the set temperature using an on/off or PID control algorithm
- Multiple syringe heater pads can be attached to the control unit
- Heated syringe can be mounted on a syringe pump
- Syringe Heater Kit includes Control-Unit power cable and Primary-Syringe-Heater-Pad
| 注射器微量泵(USA)-精确可编程序-特价 | ||||||||||||||||||||||||||||||||||||||||||||||||||||
| [ 作者:佚名 | 来源:本站原创 | 点击数:15 | 更新时间:2009-1-29 ] | ||||||||||||||||||||||||||||||||||||||||||||||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||
相关信息
昆山市超声舒美牌KQ2200DV系列清洗仪器有限公司 2010-7-28
有机合成实验室移液整体解决方案(二) 2010-7-27
有机合成实验室移液整体解决方案(二) 2010-7-27
搅拌器在有机化学实验室中的使用方法 2010-7-23
普通生物显微镜的使用方法及其注意事项 2010-7-14
显微镜基础理论与知识 2010-7-14
显微镜标本制作方法 2010-7-14
水泥颗粒形貌对其性能影响的研究 2010-7-14
CMOS的成像原理及发展 2010-7-14
CCD成像原理与分类 2010-7-14
生化培养箱 2010-7-14
如何配置PH计的氯化钾保护液 2010-7-2
超声波清洗机的广泛应用 2010-7-2
卡尔费休水分测定仪维护注意事项 2010-7-2
实验室用离心机使用时的注意事项 2010-7-2美国健康医疗仪器国际公司
地址:成都市上东大街36号.新和•春禧时代12-11
邮政编码:610016
联系人:张帆,
电话:028-86653817
传真:028-86653817
Email:ahmedical@verizon.net
展台:点击进入
邮政编码:610016
联系人:张帆,
电话:028-86653817
传真:028-86653817
Email:ahmedical@verizon.net
展台:点击进入